Ver. 1.11.00 (23 Jun 2025)
											Added 
- analog output: setting of current/voltage value corresponding to an alarm condition detected by the instrument
 
 
Changed 
- Load batching program: batched formula not shown on the displayed during the procedure
 
- Inclinometer function: improved the performances of the compensation algorithm
 
 
Fixed 
- Digital equalization: wrong reading of the coefficients by remote
 
- Digital equalization: wrong setting of equalization mode by remote
 
- Load batching program: setting of MODAFAL parameter to values greater than 9 not possible
 
- Inclinometer function: unwanted enabling of the function when accessing to the menu
 
 																					 | 
									
																		
										
											Ver. 1.10.00 (11 Mar 2025)
											Added: 
- added the DHCP menu item, to give the possibility of enabling the automatic reception of the IP configuration (only for EthernetIP Fieldbus).
 
- added the GATEWAY ENABLE menu item, to give the possibility of enabling the Gateway address set from the menu (only for EthernetIP Fieldbus).
 
 																					 | 
									
																		
										
											Ver. 1.09.08 (17 Dec 2024)
											Fixed: 
- CLM8: 16 channels mode, if load cells are connected to the slave device only, temporary error messages are displayed
 
- Reading of digital input configuration by remote doesn't return the input index
 
- Reading of digital output configuration by remote doesn't retur the output index
 
 																					 | 
									
																		
										
											Ver. 1.09.07 (25 Nov 2024)
											Added: 
- Firmware version compliant with INMETRO regulation
 
- Modbus protocol: address 0 supported for broadcast communication
 
- Analog output: output current or voltage value configured by remote
 
- RS485 serial port: continuous transmission YHL protocol available
 
- New inclinometer models supported
 
- Equalization: axis mode available
 
- Equalization: manual correction for calculated coefficients
 
- Weight value expressed in e/10 available by remote
 
- CLM8: 16 channels mode available for legal for trade applications
 
 																					 | 
									
																		
										
											Ver. 1.08.16 (14 Oct 2024)
											Internal changes without any impact on the user 																					 | 
									
																		
										
											Ver. 1.08.15 (22 Aug 2024)
											Internal changes without any impact on the user 																					 | 
									
																		
										
											Ver. 1.08.13 (18 Jul 2024)
											Fixed 
- MODBUS communication stops working after the start of a batching cycle
 
 																					 | 
									
																		
										
											Ver. 1.08.12 (11 Apr 2024)
											Fixed 
- connected loadcells not detected at the instrument startup
 
 																					 | 
									
																		
										
											Ver. 1.08.11 (06 Feb 2024)
											Fixed 
- firmware update procedure, executed with Instrument Manager, abruptly interrupted durint step 3 of 4
 
 																					 | 
									
																		
										
											Ver. 1.08.10 (02 Feb 2024)
											Fixed 
- Procedure for access to legally relevant parameters not working, if executed by remote
 
- Semi-automatic zero execution by remote: execution state not correctly updated
 
- Preset tare remote execution by remote: not allowed if the current gross weight is equal to zero
 
- ETHERNET TCP/IP interface: CONTIN string not transmitted if a properly set digital input is closed
 
 																					 | 
									
																		
										
											Ver. 1.08.07 (08 Jun 2023)
											Fixed 
- remote setting of the selection of channels to activate (via commands 6575 and 6094) did not work.
 
- remote reading of the Filter Type parameter (via command 6520) returned an incorrect value.
 
- in instruments approved for legal trade, the "reference number" was erroneously increased each time the menu was accessed.
 
 																					 | 
									
																		
										
											Ver. 1.08.06 (08 Feb 2023)
											Fixed 
- bug-fix: setting the SWITCHING OF THE ALARM/TOLERANCE RELAY ON WEIGHT batching constants from Instrument Manager, only the value was saved on the instrument but not the functionality.
 
- bug-fix: setting the SLOW batching constant from the Instrument Manager, only the value was saved on the instrument but the surrounding parameters were not adjusted.
 
- bug-fix: the "NodeId" parameter related to the Powerlink fieldbus was not set correctly.
 
- bug-fix: the reading of the mV values per channel from fieldbus was not managed correctly.
 
- bug-fix: automatic batching was interrupted in case a manual batching was performed first.
 
 																					 | 
									
																		
										
											Ver. 1.08.05 (17 Nov 2022)
											Internal changes without any impact on the user 																					 | 
									
																		
										
											Ver. 1.08.04 (09 Nov 2022)
											Fixed 
- Analog output: the output signal remains set on the gross weight even if the net weight has been selected
 
 																					 | 
									
																		
										
											Ver. 1.08.03 (05 Nov 2022)
											Added 
- Added parity menu in case of inclinometer function. Default: EVEN
 
 
Fixed 
- Sensitivity setting in theoretical equalization does not consider the divisions
 
 																					 | 
									
																		
										
											Ver. 1.08.02 (11 Oct 2022)
											Fixed 
- The ordering of the bytes in the EXCHANGE REGISTERS of the CANopen/DeviceNet protocol has been reversed.
 
- The ordering of the bytes in the EXCHANGE REGISTERS of the PROFINET-IO protocol (Single port model) has been reversed
 
 																					 | 
									
																		
										
											Ver. 1.08.01 (10 Sep 2022)
											Internal changes without any impact on the user 																					 | 
									
																		
										
											Ver. 1.08.00 (07 Sep 2022)
											Added 
- Add mV reads for each channel on fieldbus registrers
 
- Add tilt sensor protocol ACS-020-2-SC00-HE2-PM
 
 
Correzioni 
- Change Overload Managment: add singol channel error visualization, add fieldbus publication, add log event
 
- Change Load Distribution Management: add fieldbus publication, change display visualization.
 
 																					 | 
									
																		
										
											Ver. 1.07.26 (13 Jul 2022)
											Fixed 
- Command 6561: configuration wasn't maintained when the instrument is restarted
 
- ASCII protocol: fixed commands of semi-automatic zero and net
 
- Commands 105/106: error adding a new point in real calibration
 
- Command 6094: returned feedback before waiting for the command execution
 
 																					 | 
									
																		
										
											Ver. 1.07.25 (25 May 2022)
											Fixed 
- PROFIBUS-DP:
- The ordering of the bytes in the EXCHANGE REGISTERS has been reversed
 
- The COMMAND EXECUTION STATUS has been mapped in the EXCHANGE REGISTERS
 
- The READING DIVISIONS REGISTERS OF EACH CHANNEL have been published
 
 
 
 																					 | 
									
																		
										
											Ver. 1.07.24 (09 Apr 2022)
											Fixed 
- When ERCEL condition is active, status register is no more updated
 
- ETHERNET/IP: W1 exchange register byte order not correct
 
 																					 | 
									
																		
										
											Ver. 1.07.23 (24 Mar 2022)
											Fixed 
- Fixed the ordering of the words in the Exchange Register W1 in Little Endian mode in the fieldbuses that provide for the double ethernet port
 
 																					 | 
									
																		
										
											Ver. 1.07.22 (11 Feb 2022)
											Fixed 
- In the procedure for tare weight zero when the diagnostic on zero is enabled, the load divisions were also displayed for non-existent channels
 
- Fixed the visualization on the display of the word "ForP" during printing; now “Wait” displayed steadily.
 
- Fixed Endianness of the R1 / W1 exchange register in the Modbus protocol
 
- In the CC-Link BPSR procedure, the RXn0 bit was not resetted at the completion of the procedure
 
- Sending the Modbus command 6559 caused the instrument to crash at the next reboot
 
- In the batching printout, the GROSS / NET line was not printed.
 
 																					 | 
									
																		
										
											Ver. 1.07.18 (06 Jan 2022)
											Fixed 
-  Alibi memory: ID code not correctly initialized during instrument startup. It's not possible to read previously soterd weighings
 
 																					 | 
									
																		
										
											Ver. 1.07.16 (27 Nov 2021)
											Fixed 
- Reading of the load percentages on each channel by communication interface
 
- Menu sections specific for batching load program
 
 																					 | 
									
																		
										
											Ver. 1.07.15 (06 Oct 2021)
											Added 
- CC-LINK certified version by CPLA
 
- Implemented overload monitoring function for single channel
 
- Full remote control of the instrument, on all fieldbuses
 
 																					 | 
									
																		
										
											Ver. 1.06.06 (24 Feb 2021)
											Fixed 
- Channel overload function not working on all active channels
 
 																					 | 
									
																		
										
											Ver. 1.06.05 (23 Feb 2021)
											Fixed 
- TLM8, analog output: if voltage mode is set, at the successive instrument switch on, current mode is activated
 
 																					 | 
									
																		
										
											Ver. 1.06.04 (15 Jan 2021)
											Fixed 
- MODBUS-RTU protocol or MODBUS/TCP interface: setpoints 3,4 and 5 value not correctly published in the specific registers
 
 																					 | 
									
																		
										
											Ver. 1.06.03 (17 Dec 2020)
											Internal changes without any impact on the user 																					 | 
									
																		
										
											Ver. 1.07.00 (13 Nov 2020)
											Added 
- Instrument management by remote: with full access to all the configuration parameters and available functionalities
 
- Batching LOAD program available
 
- CC-LINK fieldbus: certified interface, compliant with CLPA consortium specifications
 
 																					 | 
									
																		
										
											Ver. 1.06.02 (24 Sep 2020)
											Internal changes without any impact on the user 																					 | 
									
																		
										
											Ver. 1.06.00 (22 Sep 2020)
											Added 
- Real calibration, command code 106: a new calibration point is added
 
- Real calibration, checks executed when a new calibration point is defined:
- Operation not executed if the sample weight is equal to 0
 
- Operation not executed if the sample weight is used in one of the currently defined calibration points
 
 
 
 
Changed 
- Real calibration, command code 101 original functioning restored. It deletes all previously registered calibration points and add a new one
 
 																					 | 
									
																		
										
											Ver. 1.05.26 (18 Sep 2020)
											Internal changes without any impact on the user 																					 | 
									
																		
										
											Ver. 1.05.24 (11 Sep 2020)
											Internal changes without any impact on the user 																					 | 
									
																		
										
											Ver. 1.05.22 (01 Sep 2020)
											Internal changes without any impact on the user 																					 | 
									
																		
										
											Ver. 1.05.20 (29 Jul 2020)
											Fixed 
- Test menu: distribution of load percentages about tare weight zero are displayed only if automatic diagnostic of load distribution on zero is active
 
 																					 | 
									
																		
										
											Ver. 1.05.18 (24 Jun 2020)
											Fixed 
- Zero tracking: instrument working very slow, due to unnecessary repeated data storing in the internal memory
 
 																					 | 
									
																		
										
											Ver. 1.05.16 (13 Jun 2020)
											Fixed 
- Releay outputs management: in case of setpoint mode, the output is activated during the instrument power up, when weight reading is not stable and correct yet
 
- CLM8, 16 channels system: system setup not correct
 
 																					 | 
									
																		
										
											Ver. 1.05.14 (24 Jan 2020)
											Internal changes without any impact on the user 																					 | 
									
																		
										
											Ver. 1.04.10 (23 Jan 2020)
											Internal changes without any impact on the user 																					 | 
									
																		
										
											Ver. 1.05.13 (21 Dec 2019)
											Internal changes without any impact on the user 																					 | 
									
																		
										
											Ver. 1.05.12 (29 Nov 2019)
											Fixed 
- Legal for trade instrument: reference number wrongly increased, if jumper is used for the access to legally relevant parameters
 
 																					 | 
									
																		
										
											Ver. 1.05.10 (05 Nov 2019)
											Added 
- First firmware version distributed with the service software Instrument Manager
 
 																					 | 
									
																		
										
											Ver. 1.04.09 (07 Sep 2019)
											Added 
- CONTIN protocol: ST menu item added, to select to include the stable weight condition in the transmitted string
 
 
Fixed 
- CONTIN protocol: MODT/MOD_TD selection not visible in the menu
 
 																					 | 
									
																		
										
											Ver. 1.04.08 (11 Jun 2019)
											Fixed 
- MODBUS-RTU protocol: setpoints value writing doesn't work
 
 																					 | 
									
																		
										
											Ver. 1.04.07 (16 May 2019)
											Changed 
- Converter points reading: selected mode is not stored in the internal memory. It has to be set any time the instrument is switched on
 
 
Fixed 
- RS232 serial port: delay paramter has 60ms as default value
 
 																					 | 
									
																		
										
											Ver. 1.04.05 (31 Jan 2019)
											Internal changes without any impact on the user 																					 | 
									
																		
										
											Ver. 1.04.04 (07 Nov 2018)
											Internal changes without any impact on the user 																					 | 
									
																		
										
											Ver. 1.04.03 (23 Oct 2018)
											Internal changes without any impact on the user 																					 | 
									
																		
										
											Ver. 1.04.02 (19 Sep 2018)
											Internal changes without any impact on the user 																					 | 
									
																		
										
											Ver. 1.04.01 (28 Jul 2018)
											Fixed 
- active channels raw points value reading: byte order of the exchanged data not correct
 
 																					 | 
									
																		
										
											Ver. 1.04.00 (25 Jul 2018)
											Added 
																					 | 
									
																		
										
											Ver. 1.03.16 (21 Mar 2018)
											Fixed 
- Boot mode: message 'j' displayed when instrument enters the boot mode, instead of  'b' message
 
- Fixed active channels manual selection 
 
 																					 | 
									
																		
										
											Ver. 1.03.13 (15 Nov 2017)
											Fixed 
- MODBUS-RTU protocol: setpoints registers value is wrongly equal to 0, if the reading MODBUS request doesn't start exactly with their address (40019-40024)
 
- Active channels selection: if requested by menu, a wrong number of acive channel is detected
 
 																					 | 
									
																		
										
											Ver. 1.03.11 (11 Oct 2017)
											Internal changes without any impact on the user 																					 | 
									
																		
										
											Ver. 1.03.10 (24 Aug 2017)
											Internal changes without any impact on the user 																					 | 
									
																		
										
											Ver. 1.03.09 (23 Aug 2017)
											Internal changes without any impact on the user 																					 | 
									
																		
										
											Ver. 1.03.08 (18 Jul 2017)
											Internal changes without any impact on the user 																					 | 
									
																		
										
											Ver. 1.03.07 (28 Jun 2017)
											Added 
- Info menu, added section Generic, showing the following data:
- Instrument name
 
- Software code
 
- Active program
 
- Firmware version
 
- Serial number
 
 
 
- Fieldbus interface: "BUS ER" error message introduced, to signal issues regarding the communication between internal components related to fieldbus interface section
 
 
Changed 
- Fieldbus interface: specific parameters value changing applied without instrument reboot
 
 																					 | 
									
																		
										
											Ver. 1.03.05 (19 May 2017)
											Internal changes without any impact on the user 																					 |